NMR TUTORIAL
BASIC OPERATIONS
ADVANCED OPERATIONS
The nuclear magnetic resonance spectrometer (usually abbreviated NMR) has three basic parts:

NMR Sample
Preparation
You cannot just put a pure liquid or solid into the NMR. There are a series of steps you need to go through to prepare your sample.
Deuterated solvents are very expensive so try not to waste them.
The NMR computer is fairly easy to use—the Delta software that runs everything works very similarly to a Windows program. Usually when you get on the NMR computer there will be four windows open.
DO NOT CLOSE THESE WINDOWS!!!
If an NMR spectrum is open you can close it.
There are four things you have to do to obtain an NMR spectrum on your sample:

SAMPLE CONTROL WINDOW
Load the Sample

Click ONCE on the
to eject the
previous sample. When you do this, the
will empty and the
green bar will disappear and the red bar will appear. At the same time, the
Spinner icon
will fall down and the
will be lit. You will see
the sample tube pop up in the magnet (it’s pushed up by a cushion of air).
DO
When you place the NMR tube into the spinner, make sure the spinner is pulled all the way apart. Use the depth gauge to position the tube to the right height in the spinner. Try not to touch the bottom of the spinner assembly or the bottom of your NMR tube.

Place the sample/spinner assembly into the magnet where it
will float on the cushion of air. Go back to the computer and click ONCE on the
icon to load the
sample. The
will fill blue again
and the
will hop up.
After you have loaded the sample and it is spinning, you need to lock the instrument on the correct solvent. Go to the Solvent area and scroll to the correct solvent and click on it. Usually, we will be using CHLOROFORM-D (CDCl3).

After you have selected the solvent, you must lock the
spectrometer on it by going to the Lock Control area and clicking ONCE on autolock
. You will see the red bars below the autolock turn yellow:

When the instrument has locked on the solvent, these bars will turn green:

After the instrument has locked, you need to shim the magnet so that the sample gets the maximum irradiation. Go over to the shim control box next to the monitor and press the Z1-Z2 button. Then turn the Z1 knob either clockwise or counterclockwise (it doesn’t matter which way you turn it to start). As you turn the knob you will hear a clicking sound. You are looking for the biggest lock signal number possible. If you turn the Z1 knob one way and the lock signal goes down, then you need to turn the knob the other way. Don’t turn the knob too fast or slow, just a nice steady motion. When you have gotten the number as high as possible with the Z1 knob, switch to the Z2 knob and repeat the process. There is no magic number to shoot for, just the biggest you can get. If nothing seems to be happening, look at the sheet posted with the latest shim numbers and turn the knobs to those numbers and shim from there. Do not touch the Z3-Z4 button.

SHIM BOX
Locate the experiment window. If you are running a 1H
NMR look for single_pulse.exp and if
you are doing 13C look for single_pulse_dec.exp.
The experiment window is divided into sections by color. One software quirk
is that the cursor arrow must be in the box in order to change the value.
In the green section, change the filename to something that you can find later. Use a different filename for each sample (don’t use your name for every sample). DO NOT USE A BACKSLASH (/) IN YOUR FILENAME, JUST NUMBERS, LETTERS, AND UNDERSCORES. The computer will automatically save each spectrum you run. Click once on the auto_gain box to put a red check mark in it (this may already be checked from the previous person).

TOP OF THE EXPERIMENT WINDOW
In the pink section, make sure the solvent is the same as the one in the sample control window. Do not touch any other things in this section.

The purple section controls such things as the width of your spectrum and how many scans you are going to run. You usually only need to run 4 scans on the sample (50 for 13C) and the default sweep width is usually OK. The default 1H spectrum is centered at 5 ppm and runs 7.5 ppm to either side (a total width of 15 ppm). If you suspect that you have a signal outside that range, then set the sweep width accordingly. Don’t change any of the other values. Do not change any values inside the yellow section.

Once you have changed the filename and other values, click ONCE on
. You will hear Jill ask you to check your filename and a box
will appear:

Click on the
button and the
experiment will start running. You can check to make sure your experiment is
running by going to the spectrometer control window. You should see one
experiment listed with an asterisk (*) by it. The asterisk indicates that
experiment is running. If there are more than one experiment listed than you or
the person before you hit
more than once. Just highlight the experiment
that you want to delete from the queue and click
. Some other boxes will come up, just answer OK to them. You
can highlight all the experiments and stop them all and then just go back and
submit your experiment again.

SPECTROMETER CONTROL WINDOW SHOWING TWO EXPERIMENTS IN THE QUEUE
You can follow the course of the experiment in the Delta window. It should look something like this:

DELTA WINDOW SHOWING AN EXPERIMENT RUNNING
If the window does not look like this, there is a problem with the instrument and you should consult an instructor. Your spectrum should come up in a few minutes, if not, ask an instructor.
Manipulating the Spectrum
When the spectrum comes up, it will not completely fill the screen. You should maximize it by clicking on the square in the upper right corner (just like any normal windows program). The window will be called 1D Processor: your filename. Below is a typical 1H NMR spectrum of ethylbenzene (the filename is ethyl and it’s version 1) :

INITIAL APPEARANCE OF ETHYL.1
You will not use the majority of buttons on the screen. The
first thing you should do is to zoom in on the portion of the spectrum you are
interested in, in the case of ethyl.1,
from ~8 ppm to ~0 ppm. To zoom, you need the zoom tool bar which you can access
by hitting the F1 key on the
keyboard. If the tool bar does not appear, click on the
which is in the
upper right of the spectrum. The tool bar will appear in the upper right of the
spectrum as shown below. The default tool is the zoom tool,
(the other buttons do
things you shouldn’t need). Move the cursor into the spectrum, click and hold
the left mouse button to start the zoom rectangle, size the rectangle by moving
the mouse, and release the mouse button to complete the zoom. If you make a
mistake, just hit the Backspace key
on the keyboard. Sometimes, solvent signals are much larger than the signals
from your sample. In those cases, you don’t have to zoom on the top part of the
solvent signal, just zoom to the height of your highest sample signal. If you
need to get back to your original spectrum, hit the Home key on the keyboard.
Once you have zoomed on the correct portion of your spectrum, you need to integrate your signals.

ETHYL.1 WITH ZOOM TOOLBAR AND ZOOM TOOL SELECTED
If you are running a 13C NMR spectrum, you will not need to integrate the spectrum. You should always integrate a 1H NMR spectrum. To access the integration tool bar, hit F8 on the keyboard. The integration tool bar has several keys you need.
![]()
Is the tool used to
set the integrals. When you integrate a signal you are
measuring the area of that signal, so the integral must start to the left of
the signal and end to the right of it. To set an integral, move the cursor to
the gray area below the spectrum. Click and hold the left mouse button
somewhere to the left of the signal you are integrating. A small blue square
will appear. Move the cursor to the right and you will see a second blue square
appear that is connected to the first by a blue line. The second square will
move along the green line of the spectrum. A yellow integral line will appear
as you integrate the signal. Keep moving until you reach the end of the signal
and release the mouse button. Repeat this process for the other signals you
want to integrate. Red integral lines will be present for signals you have
already integrated. The computer will assign integral values that will seem
meaningless to you until you normalize the values (see
below). The computer
arbitrarily thinks the largest signal should have an area of 1 and scales the
others accordingly. You will learn that this practice does not reflect reality.
If you know a signal is due to the solvent or some other impurity, don’t
integrate it.

ETHYL.1 SHOWING TWO INTEGRALS, BEFORE NORMALIZATION
Is the tool used to
delete any integrals that you don’t need or if you made a mistake on a signal.
You must click on the red or yellow integral line, not the signal itself or the
integral value.
Is the tool for
selecting an integral. You must click on the integral
line, not the signal or integral value. When you need to normalize your
integral values, click on an integral line that you are fairly confident you
know the value of. For instance, if you know a signal comes from a CH3
group, then you know its integration should be 3. When you select an integral,
its line and number will turn yellow. Press the Ctrl and N keys
simultaneously and you will get the Integral
Normal Value window:

The default value is 1. Change the value to what you think the integration for that signal should be. Remember to move the mouse cursor into the window. If your integral values are not close to whole numbers, try to normalize a different integral or try a different number. If you see an M at the end of the integral value it means that number is multiplied by 0.001. So, 84.5M is really 0.0845. If one of your integral areas is this low, it’s probably an impurity since you must have a whole number of protons in a molecule.
You are now almost ready to print. If you are happy with
what appears on the screen, click ONCE on the
button at the top of
the window. If you need to zoom in on a different portion of the spectrum than
what is displayed, go back to the zoom tool and do it. The integrals will
vanish. You can get them back by hitting F8 before you print.
When your spectrum has printed, close [
]the spectrum (don’t minimize it!).
Go back to the Sample Control window and eject
your sample. You
must always leave a sample loaded in the magnet! If someone else needs to
run a sample after you, just hand them the spinner. If you are the last to use
the instrument for awhile, load the tube with the yellow cap. Be sure you load
it! Don’t just leave it floating over the magnet. Take your sample back to the
lab, discard the liquid in the halogentated waste container, clean it using the
NMR tube cleaner, and put the tube in the oven.
ADVANCED OPERATIONS
If you need to see an old spectrum again for some reason,
it’s easy to get to it. Go to the Delta window and click and hold on
Processors. Scroll to Process 1D and let go the mouse button. A blank spectrum
window will appear. Click once on
and you should get a
window with all the filenames. Scroll down to yours (the ALL CAPS filenames are
listed first) and highlight it. To the right you will see a series of numbers
which are the different versions of that file. Highlight the highest number
first and click OK. Your spectrum should appear, although it takes some time.
If you get an FID (see below) instead of a spectrum, you need to process the
FID. To do that, click once on
and you will get a new
window. You are looking for the process list called std_proton_autophase.list if you are processing a 1H spectrum
or std_carbon_autophase.list for 13C.
Highlight the process list and click OK. That window will disappear and you’ll
be back in the spectrum. Click once on
and wait a minute for
the spectrum to appear.

AN UNPROCESSED FID
Hopefully you won’t have to do anything outlined below, but just in case…
What
if the experiment window isn’t there?
If no experiment window is present (or if the wrong one is
there), you need to open one. First make sure it hasn’t been minimized.
Otherwise, go to the Spectrometer Control window and click on
. You will then get a new window that looks like:

If the experiment you need isn’t listed, click ONCE on
and you will get a new
window with a list of experiments. Scroll down to the one you need (single_pulse.exp for 1H or single_pulse_dec.exp for 13C)
and highlight it. Click the OK button and the experiment window will come up.

What
if the Sample Control Window isn’t there?
Always make sure the window you need hasn’t been minimized.
If the Sample Control Window has been closed, go to the Spectrometer Control
window and click ONCE on
and the Sample Control
Window should come up.
What
if the Spectrometer Control Window isn’t there?
If the Spectrometer Control Window has been closed
inadvertently, go to the Delta Window and click ONCE on
and the Spectrometer
Control Window should appear. You will need to connect to the spectrometer by
highlighting nmr2-FREE-Centre College and clicking ONCE on
.
What if the sample isn’t
spinning?
You will know the sample
isn’t spinning if the Current value
is 0[Hz] or far off from the Target value of 15[Hz]. If the sample isn’t spinning, eject it first. Make sure the
spinner is pulled all the way apart. If it is, take out the NMR tube and wipe
the bottom of the wide part of the spinner with a kimwipe. Put the spinner back
together, re-insert the tube, and load the sample again. If it still doesn’t
spin, consult an instructor.

What if the experiment is not running?
The usual reason the experiment won’t be running is because
you forgot to hit OK when Jill asked you to check your sample ID. Go to the
Spectrometer Control Window to see if there is an experiment in the queue. If
not, go back to the Experiment Window and click ONCE on
. If there is an experiment listed in the queue, but no * by
it, highlight the experiment and click ONCE on OK. It should start running. If
that doesn’t work, consult an instructor.